title: "Shimadzu AGX-V Precision Universal Testing Machine"
space: "Factory Docs"
url: "https://factory.uga.edu/docs/equipment/shimadzu-agx-v"
updated: "2026-11-19"
---

## Overview

The Shimadzu AUTOGRAPH AGX-V is a precision universal testing machine used for controlled tensile, compression, bending, and related mechanical tests. The laboratory system is installed with a dedicated, non-networked Windows computer running TRAPEZIUM-X software.

The lab has used the AGX-V for pecan-shell cracking research, including tests that vary humidity, temperature, applied force, loading direction, and compression-plate design.

> **Local workstation:** The attached Windows computer is intentionally non-networked. Sign in to TRAPEZIUM-X with username `admin` and password `admin`. These are application credentials, not necessarily the Windows account credentials.

> **Configuration warning:** AGX-V systems are available in multiple frame capacities and can use different load cells, grips, fixtures, and extensometers. Before testing, record the capacity and serial number of the installed frame, load cell, and accessories. Never infer the safe test force from the AGX-V product-family maximum.

---

## System Components

A typical AGX-V installation includes:

- AGX-V load frame and crosshead
- load cell
- upper and lower grips or compression fixtures
- smart controller
- emergency-stop control
- adjustable upper and lower stroke limit switches
- dedicated Windows computer
- TRAPEZIUM-X test software
- optional extensometer or other measurement channels

The measurement chain is:

```text
Specimen
   |
Grip / Fixture
   |
Load Cell ---------> Force
   |
Crosshead Encoder -> Position / Displacement
   |
Extensometer ------> Specimen Strain, when installed
   |
AGX-V Controller
   |
TRAPEZIUM-X
   |
Raw Data + Calculated Results + Report
```

---

## Capabilities

Depending on the installed fixtures and sensors, the AGX-V can perform:

- tensile tests
- compression tests
- three-point or four-point bending tests
- puncture tests
- peel and adhesion tests
- cyclic or repeated-load tests within system limits
- load-hold and displacement-hold sequences
- strain-controlled testing with a compatible extensometer

The AGX-V control system supports approximately 1 kHz response control and data acquisition up to 10 kHz. A high sample rate is useful for abrupt fracture events, but it also increases output size and does not correct poor fixturing or an unsuitable sensor range.

---

## TRAPEZIUM-X Workstation

### Local access

The software is installed on the computer physically attached to the tester.

```text
TRAPEZIUM-X username: admin
TRAPEZIUM-X password: admin
```

Because the workstation is non-networked:

- Do not expect cloud or network storage to be available.
- Do not connect it to Wi-Fi or Ethernet without laboratory authorization.
- Transfer results only through the lab-approved removable-media process.
- Scan removable media on a managed, networked computer before and after use.
- Do not install software, drivers, or Windows updates without confirming compatibility with the instrument.
- Preserve the working instrument configuration before making changes.

### Data handling

Use a predictable directory structure on the local computer:

```text
AGX-V_Data/
└── YYYY/
    └── Project_Name/
        └── YYYY-MM-DD_Test_Series/
            ├── Methods/
            ├── Raw/
            ├── Exports/
            └── Reports/
```

Recommended specimen naming:

```text
<Project>_<Material>_<Condition>_<SpecimenID>_<Date>
```

Example:

```text
PECAN_Desirable_65RH_P012_2026-11-19
```

Export both the native TRAPEZIUM-X result and an open tabular format such as CSV. A PDF report is useful for review but is not a substitute for raw data.

---

## Before Testing

### 1. Identify the method

Record:

- applicable test standard or internal procedure
- material and specimen geometry
- expected maximum load
- desired control mode
- test speed or strain rate
- preload
- termination criteria
- required output values

Common standards include:

- ASTM D638 for tensile properties of plastics
- ASTM E8/E8M for tension testing of metallic materials
- ASTM D3039/D3039M for polymer-matrix composites
- ASTM D790 for flexural properties of plastics
- ASTM E9 for compression testing of metallic materials
- ISO 527 for tensile properties of plastics
- ISO 6892-1 for metallic tensile testing

Use the standard applicable to the material and research question. Do not copy a convenient test speed from an unrelated method.

### 2. Select the load cell

Choose a calibrated load cell with sufficient capacity while keeping expected measurements in a useful portion of its range.

A practical target is to keep the expected peak force roughly above 10% of load-cell capacity when possible, without risking overload. For example, a specimen expected to fail near 1 kN is usually measured more effectively with a 5 kN or 10 kN load cell than with a 100 kN load cell.

Never exceed the rated capacity of the load cell, grips, fixtures, pins, or frame. The safe system limit is the **lowest capacity in the complete load path**.

### 3. Select and inspect fixtures

Confirm that fixtures are appropriate for:

- tension, compression, or bending
- expected force
- specimen geometry
- temperature/environment
- surface condition

Inspect for:

- loose fasteners
- damaged grip faces
- worn pins
- fixture misalignment
- contamination
- previous overload damage

### 4. Inspect the specimen

Record relevant dimensions and conditioning information before mounting. Depending on the method, this may include:

- width
- thickness
- gauge length
- cross-sectional area
- mass
- humidity conditioning
- temperature conditioning
- material orientation
- visible defects

Use a calibrated micrometer or caliper appropriate to the specimen and standard.

---

## Startup Procedure

1. Verify the test area is clear.
2. Confirm the installed load cell and fixtures match the planned method.
3. Inspect cables, grips, and crosshead area.
4. Confirm the emergency stop is released.
5. Check the physical upper and lower stroke-limit switches.
6. Power on the AGX-V system using the established lab sequence.
7. Start the attached Windows computer.
8. Open TRAPEZIUM-X.
9. Sign in with `admin` / `admin`.
10. Confirm communication with the tester.
11. Allow the load cell and electronics to warm up if required by the test procedure or calibration documentation.
12. Verify force and position readings before installing the specimen.

Do not begin a test if the displayed force is unstable, the frame reports an error, or the active hardware configuration does not match the installed hardware.

---

## Crosshead and Stroke Limits

The AGX-V has an intelligent crosshead and physical stroke-limit switches. Set the upper and lower limits before jogging near fixtures or installing a specimen.

The limit switches are damage-mitigation devices, not a substitute for a correct test method.

Before motion, consider:

```text
Available travel
  - fixture height
  - specimen length
  - expected extension/compression
  - stopping distance
  - safety clearance
= usable test stroke
```

Jog at low speed when fixtures are close together. Keep hands out of the load string during crosshead motion.

---

## Creating or Loading a Test Method

In TRAPEZIUM-X, load a validated existing method when one applies. Otherwise create a method using the guided condition-setting workflow.

A method should define at least:

- test type
- control channel
- preload
- crosshead speed or strain rate
- sampling conditions
- specimen dimensions
- calculation items
- break or stop detection
- force limit
- displacement limit
- return behavior
- report layout

### Method control example

```text
1. Apply small preload
2. Zero or establish extension reference
3. Load at specified crosshead speed
4. Record force, displacement, and strain
5. Detect fracture or reach termination limit
6. Stop crosshead
7. Return only when safe
8. Save native result and exports
```

Validate a new method using a sacrificial or representative specimen at reduced speed before running the full series.

---

## Mounting a Tensile Specimen

1. Confirm the crosshead is stopped.
2. Open the grips sufficiently for the specimen.
3. Install the specimen in one grip.
4. Align the specimen with the load axis.
5. Close the first grip using appropriate clamping force.
6. Install the specimen in the second grip without adding unintended preload.
7. Verify that the specimen is centered and not twisted.
8. Confirm adequate clearance for grips, extensometer, and expected elongation.
9. Apply the method-defined preload if required.
10. Attach the extensometer only according to its procedure and travel limits.

Poor alignment introduces bending and can invalidate tensile results. Slippage, grip-induced fracture, or fracture outside the acceptable gauge region should be recorded and handled according to the applicable standard.

---

## Running a Test

Before selecting **Start**, confirm:

- the correct method is active
- specimen ID and dimensions are correct
- load-cell selection is correct
- force has been balanced/zeroed as required
- displacement or extension reference is correct
- extensometer limits are sufficient
- force and stroke limits are set
- the enclosure or safety shield is positioned when fragments may be ejected
- no person is inside the hazard area

During the test:

- Observe from outside the pinch and fragment zones.
- Do not touch the frame, specimen, grips, or extensometer.
- Monitor force and displacement for unexpected behavior.
- Stop the test if the specimen slips, fixtures loosen, or the trajectory is inconsistent with the method.

Use the emergency stop for an imminent hazard, not as the routine method for ending a normal test.

---

## Stress, Strain, and Extension

### Engineering stress

```text
σ = F / A₀
```

where:

- `σ` is engineering stress
- `F` is axial force
- `A₀` is original cross-sectional area

### Engineering strain

```text
ε = ΔL / L₀
```

where:

- `ε` is engineering strain
- `ΔL` is change in gauge length
- `L₀` is original gauge length

Crosshead displacement is not automatically equal to specimen extension. It includes compliance from the frame, grips, fixtures, and possible specimen seating or slip.

Use an extensometer when the method requires accurate strain, elastic modulus, yield strain, or strain-rate control.

---

## Tensile-Test Quality Checks

Review each result for:

- specimen break location
- grip slip
- preload behavior
- force-zero stability
- curve discontinuities
- extensometer removal or travel limit
- unexpected compliance
- incorrect specimen dimensions
- sensor saturation
- premature termination

A smooth-looking curve is not sufficient evidence of a valid test. Check whether the physical failure mode and test setup satisfy the governing method.

---

## Pecan Compression and Cracking Tests

For pecan-shell research, compression tests can be used to characterize cracking force and energy under controlled conditions.

Relevant experimental variables include:

- cultivar
- nut orientation
- moisture or humidity conditioning
- temperature
- plate geometry
- loading direction
- crosshead speed
- preload/contact threshold

A useful test record includes:

```text
Specimen ID
Cultivar
Conditioning temperature
Conditioning relative humidity
Conditioning duration
Mass
Orientation
Compression plate ID
Crosshead speed
Peak force
Displacement at first crack
Energy to first crack
Observed failure mode
```

For brittle cracking events, use a sampling rate high enough to resolve the force drop. Confirm that the selected load cell will not be overloaded by peak impact-like forces.

Do not interpret crosshead displacement directly as shell deformation without accounting for fixture and machine compliance.

---

## Data Export and Backup

After each test series:

1. Save the native TRAPEZIUM-X project/result.
2. Export raw channels to CSV or another open tabular format.
3. Export the reviewed report to PDF if required.
4. Copy the complete test-series directory to approved removable media.
5. Transfer the data to the approved research storage location using a managed computer.
6. Verify the copied files before deleting any local data.
7. Keep at least two verified copies of irreplaceable raw data.

Include the method file with the data. Force-displacement results without the control and calculation settings are not fully reproducible.

---

## Shutdown Procedure

1. Confirm the test is complete and data are saved.
2. Remove the specimen and extensometer safely.
3. Leave adequate clearance between fixtures.
4. Return the crosshead to the lab's designated resting position if required.
5. Close TRAPEZIUM-X.
6. Shut down Windows normally.
7. Power down the controller and frame using the established lab sequence.
8. Clean fixture surfaces and remove debris.
9. Record faults, overloads, damaged specimens, or unusual behavior in the equipment log.

---

## Safety

The AGX-V can generate forces sufficient to crush fingers, eject specimen fragments, damage fixtures, or overload sensors.

Primary hazards include:

- moving-crosshead pinch and crush points
- stored elastic energy in specimens
- brittle-fracture projectiles
- grip or fixture failure
- extensometer overtravel
- load-cell overload
- unexpected crosshead return
- electrical hazards inside the frame and controller

### Required practices

- Wear eye protection appropriate to the specimen hazard.
- Use a shield or enclosure for brittle or fragmenting specimens.
- Keep hands clear whenever the crosshead can move.
- Never adjust a specimen under load.
- Never exceed the rating of the weakest load-path component.
- Set physical stroke limits before testing.
- Use tools rather than fingers to remove sharp fragments.
- Do not bypass interlocks or protective functions.
- Do not run unattended unless the procedure has been specifically approved for unattended operation.

### Emergency stop

Use the emergency stop if continued motion presents immediate risk to a person or the equipment. After an emergency stop:

1. Identify the cause.
2. Remove stored mechanical energy safely.
3. Inspect the specimen, fixtures, load cell, and frame.
4. Reset only when the hazard has been corrected.
5. Document the incident if equipment may have been overloaded or damaged.

---

## Calibration and Verification

Do not use the AGX-V for reportable quantitative work if the relevant calibration is expired.

Check calibration status for:

- load cell
- extensometer
- displacement measurement, when required
- dimensional measurement tools

Relevant standards include:

- ASTM E4 for force verification of testing machines
- ISO 7500-1 for static uniaxial testing-machine force verification
- ASTM E83 for extensometer verification and classification
- ASTM E1012 for alignment verification of tension and compression machines

A force calibration does not prove specimen alignment. Alignment matters especially for metals, composites, and other materials sensitive to bending strain.

---

## Troubleshooting

### TRAPEZIUM-X cannot communicate with the tester

Check:

- AGX-V controller power
- instrument/computer cable connections
- selected instrument configuration
- Windows device state
- whether another TRAPEZIUM-X process is open
- controller or software error messages

Because the workstation is intentionally offline, network troubleshooting is normally irrelevant to the instrument connection.

### Force reading does not return near zero

Check:

- installed fixtures and preload
- whether the load cell was balanced with the correct hardware installed
- cable condition
- temperature stabilization
- mechanical contact between fixtures
- possible load-cell overload or damage

Do not repeatedly re-zero a drifting or unstable load cell without identifying the cause.

### Specimen slips in the grips

Check:

- grip-face type
- clamping force
- specimen surface
- specimen width and thickness
- jaw condition
- alignment

Increasing grip force indiscriminately can cause grip-induced fracture or specimen damage.

### Specimen breaks at the grip

Possible causes include:

- excessive clamping pressure
- sharp grip-face transitions
- poor alignment
- inappropriate specimen geometry
- material defects
- an incorrect gripping method

Apply the validity rules in the governing standard before including the result.

### Crosshead stops unexpectedly

Check:

- force limit
- displacement limit
- physical stroke limit
- break detection
- overload protection
- emergency-stop state
- controller alarms

### Results differ substantially between specimens

Separate material variability from test-system variability. Review:

- specimen dimensions
- conditioning
- orientation
- alignment
- grip slip
- test speed
- calculation settings
- load-cell range
- fracture mode

---

## References

- Shimadzu, *AUTOGRAPH AGX-V Series Precision Universal Testing Machines*, C224-E112.
- Shimadzu, *TRAPEZIUM-X / TRAPEZIUM X-V User and Reference Documentation*.
- ASTM E4, *Standard Practices for Force Calibration and Verification of Testing Machines*.
- ASTM E83, *Standard Practice for Verification and Classification of Extensometer Systems*.
- ASTM E1012, *Standard Practice for Verification of Testing Frame and Specimen Alignment Under Tensile and Compressive Axial Force Application*.
- ISO 7500-1, *Metallic materials — Calibration and verification of static uniaxial testing machines*.